TESTED Shimadzu GC-2014AFsc GC Gas Chromat FID AOC-20i LabSolutions PC & License
TESTED Shimadzu GC-2014AFsc GC Gas Chromat FID AOC-20i LabSolutions PC & License
Couldn't load pickup availability
TESTED Shimadzu GC-2014AFsc Gas Chromatograph w/ AOC-20i, FID, Computer & Key
Sale Includes the Following Items
1x) Shimadzu GC-2014AFsc Gas Chromatograph
1x) Shimadzu AOC-20i Auto Injector w/ 10uL Syringe and 6-Sample (Short) Rack
1x) HP EliteDesk Computer w/ Shimadzu LabSolutions software pre-installed
1x) Shimadzu GC/LC License Key
1x) Includes the currently-installed Zebron ZB-1 column
Testing Included the Following:
The system was connected to Shimadzu LabSolutions and tested using nitrogen carrier gas with the installed 100 m × 0.25 mm ID × 0.50 µm capillary column. Communication with LabSolutions was confirmed, a method was created and downloaded, and the GC completed a system check successfully. The AOC-20i autosampler was recognized and completed an automated 1.0 µL liquid injection. The SPL inlet reached and maintained 180 °C, the column oven completed the programmed 50-100 °C temperature ramp, and the SFID detector reached and maintained 200 °C. Nitrogen carrier pressure and flow remained stable, the FID ignited and remained lit, and the system reached ready status and completed a full run. LabSolutions recorded a chromatogram with multiple resolved peaks from an ethanol/hexane test mixture, confirming data acquisition, detector response, oven programming, carrier control, and autosampler operation. Only the items shown are included.
Used, see photos
-BJ
Share